TMS
5000: Long Standoff Non-Contact Thickness Gage
Up to 24 inch standoff
2 (two) 30mm diameter gage heads per measurement
The TMS 5000 is a non-contact, ultrasonic thickness
measurement system for use in hostile environments where there
is airborne contamination, hot product, large environmental
temperature changes and the potential for mechanical damage to
the sensors. In addition, it is for applications requiring
0.001" accuracy or better. The TMS 5000 thickness system is
similar to the older, discontinued TMS-1000 but with the
double reference transducer and controller. The picture below
shows the older single reference bar transducer (left) with
the new, improved, patented double reference bar transducer
(right).

The TMS-5000, measures two known points in space so that if
the wave form changes it won’t affect the measurement. All
drift, for any reason, is eliminated. The TMS-5000 is intended
for environments with moderate to heavy airborne
contamination.
Ultrasonic Arrays thickness gages are used to measure
materials that are produced in a process environment. UA
thickness gages are "embedded" in the process line and measure
the product thickness as it is transferred. The gages
automatically recalibrate themselves for anything that affects
the measurement, even if the process line or fixture moves.
The result is a repeatable, accurate, real time measurement
that is more accurate and repeatable than a hand micrometer
measurement. Measurement is real time, so set up is faster and
the process can be automatically adjusted. Every product is
measured, assuring the customer's specification will be met.
Measuremate software provides sophisticated grading rules and
is used to determine the cause of variation, earlier in the
process. Measuremate is also used to optimize the production
process and provides important SPC data to control the
process.
UAI has been installing non-contact ultrasonic thickness gages
for over 20 years. Constant development has resulted in
increased accuracy, added features to allow for less than
perfect fixturing and use in ever harsher environments,
automatic calibration for everything that affects the
measurement, fail safe features, diagnostics and software that
improves quality and increases profit. These developments and
improvements have made UAI non-contact thickness gages the
standard in many industries.
UAI gages can be used as stand alone systems and do not
require a computer. Individual gages are microprocessor
controlled and are programmed via a keypad and display.
TMS-5000
Specifications
PERFORMANCE
Accuracy: +/-0.00025" or +/-0.025% of the sum of the
distances from the transducers to the target surfaces;
whichever is greater for distances up to 5" +/-0.033% from
1.5" up to 4" and +/-0.05% thereafter.
Repeatability: +/-0.05% of the distance from the transducers
to the target
Inspection Rate: Variable to 58 measurements per second
Operating Range: 0.2" - 23.999" or 5mm - 609mm (Transducer to
target)
Thickness Range:
0.004" - 9.9999" in 0.0001 resolution mode
0.004" - 99.999" in 0.001 resolution mode
0.1mm - 99.999mm in 0.001mm resolution mode
0.1mm - 999.99mm in 0.01mm resolution mode
Off-Axis Alignment: +/- 2.5 to 5 degrees depending on target
size, material and range
ELECTRICAL
Input Power:
100 VAC, 50/60 Hz
120 VAC, 60 Hz
230 VAC, 50 Hz
Maximum Power Consumption: 35 watts
Standard Interfaces: Full duplex asynchronous RS-232C, 3-wire
or RS-422A, 5-wire communications interface
Baud Settings: 300, 600, 1200, 2400, 4800, 9600, 19200, none
Word Length: 8 bit; 7 ASCII + 1 Parity Bit
Parity: unused
Stop Bit: 1 stop bit
Line Status: disable
16 bit parallel binary output, TTL Compatible 3mA Source, 20mA
Sink
Optional Interface:
Analog: 0-10 VDC or 4-20mA (up to 2500 Ohm load for 4-20mA)
Selectable analog span: 00.00" to 99.9999" or 0mm to 999.9mm
Incremental steps of voltage or current within span: 4096
steps
MECHANICAL/ENVIRONMENTAL
Dimensions and Weight:
Transducer: 5.5"H x 1.5"D .33lbs
Controller: 12"H x 12"W x 7.5"D, 25 lbs Enclosure NEMA 13
Rated
Industrial Air Hood: 11"H x 4.5"W x 4.0"D 5 lbs
Air Fan/Filter: 110V, 15"H x 14"W x 15"D 20 lbs
Temperature: Operating -20 to +50 C
Noise Immunity:
Acoustic: Filtered to eliminate noise spectrum of common
industrial environments
Electrical: Shielded for RF/EMI
Humidity: 95% maximum Non-condensing
Transducer to Controller Separation:
Standard: 10'-30' Optional: 500'
PROGRAMMABLE TRANSDUCER CONTROLLER
Display: 5 digit LED display on controller
Programming: 2 momentary push button switches on controller
provide local programming of parameters. Controller can also
be programmed remotely via the RS-232C, RS-422A interface
Signal Compensation: Continuous automatic calibration for
temperature changes
Sample Averaging: Programmable from 1 to 4095 samples per
measurement
Amplitude Filter: Programmable to filter measurements from
vibrating or poorly aligned targets
Run Mode: Programmable for free running or sample on command
by switch or contact closure
Relay Configuration: Programmable relay closure in or out of
limit set points. Electromechanical SPDT rated 2A maximum, 60
VDC
Relay Thickness Limit Set Points: Programmable high and low
relay limit set points for selectable closure, normally-in or
normally-out of limits in 0.01 inch increments
Delay to Relay Closure: Programmable relay delay from 0.002 to
65 seconds in 0.002 second increments
Relay Closure Interval: Programmable relay closure interval
from 0.002 to 65 seconds in 0.002 second increments